Clave clean Autoclave cleaning solution

Solución para limpieza y mantenimiento de autoclaves. Envase 250ml.

Instructions Safety data sheet

 

SKU: SLA250 Category

Description

Product for cleaning and maintenance of autoclave chambers.

It helps to remove debris and oxides and prolongs the life of the autoclave, dries quickly and is biodegradable.

COMPOSITION:

  • TENSIOACTIVE
  • CITRIC ACID
  • ISOPROPYL ALCOHOL
  • EXICIPIENTS

INSTRUCTIONS FOR USE:

Autoclaves without a water tank: thoroughly clean the sterilisation chamber with a sponge soaked in Clave Clean. Rinse with water. Perform a complete sterilisation cycle. Autoclaves with water tank (Class A).
1. Empty and clean the water tank, check that the injectors are not clogged.
2. Put about 30-50ml of cleaning solution in the tank.
3. Perform a vacuum sterilisation cycle to clean the system.
4. Empty the water tank of any remaining solution and refill it with distilled water.
5. Clean the sterilisation chamber with a sponge soaked in Clave Clean, then rinse thoroughly with a cloth soaked in distilled water. If you wish, you can run a distilled water cycle to rinse off any residual cleaner.
After this, the autoclave is clean and free of residues that could deteriorate it.

PRECAUTIONS

The product is intended for use in stainless steel chambers. If your autoclave is made of other materials, you should test in each case. The use must be done in a ventilated room.

CONTENTS:

250ml bottle.

Reviews

There are no reviews yet.

Only logged in customers who have purchased this product may leave a review.